28m3 NE60519-2.1993 Walk In Environmental Test Chamber APP Mobility Management
Contact Person : Jack He
Phone Number : +86-13842409825
WhatsApp : +8613842409825
Minimum Order Quantity : | 1 | Price : | $120000 >=1 Set |
---|---|---|---|
Delivery Time : | 35 days | Payment Terms : | Western Union, T/T, L/C, D/A, D/P |
Supply Ability : | 5 sets per month |
Place of Origin: | Guangdong, China | Brand Name: | Envsin |
---|---|---|---|
Certification: | CE | Model Number: | ECZ |
Detail Information |
|||
Certification: | CE | Power: | AC 380V±10%, 50Hz, 3/N/PE |
---|---|---|---|
Humidity Range: | 10%~98% RH And Other Customized | Temperature Range: | -60 ℃~+180 ℃ |
Warranty: | 12 Months | Heating Ra≤3°C/minte: | <=3 ℃/s |
Cooling Rate: | <=1 ℃/s | Test Volume: | 8m3,17m3 ,22m3 And Other Customized |
High Light: | 30000n/m2 Walk In Environmental Test Chamber,17m3 Walk In Environmental Test Chamber,17m3 temperature humidity test chamber |
Product Description
Walk In Temperature and Humidity Test Chamber With Computer Precision Control
The Walk-in chamber is a full range of reliable test chambers and the perfect test equipment for everything from testing against humidity, temperature and various climate conditions. It provides supreme temperature testing for large components and equipment. Whether you are storing many small products on carts, creating a controlled workspace for research and design, or if you need to drive large vehicles in for testing, we can design customized machine for you.
Ø Special anti-condensation function to protect specimens during climate testing. |
Ø Intelligent and high-performance refrigerating control technology: rapid heating,cooling rate. |
Ø Automatic reminder of equipment maintenance plans and software of recording malfunction. |
Ø Unique operation mode: make testing area's temperature back to room temperature after tests are finished |
Specifications
Climate Chambers | ECZ 8/010 | ECZ 17/010 | ECZ 28/010 |
Test Space Volume (m3) | 8 | 17 | 28 |
Test Space Dimension (mm) | 2000W*2000D*2056H | 2250W*2500D*3032H | 2250W*2500D*5000H |
External Dimension (mm) | 2240W*2240D*3800H | 2490W*2740D*4900H | 2490W*2740D*6860H |
Floor Load (N/m2) | 30000 | ||
ETC:-60,-40/+80,+100,+150,+180 | |||
Temp Range (℃) (No Load) | ECZ: 0/+80,+100 | ||
ETC: -60,-40/+80,+100,+150 | |||
Humid. Range( %RH) | 15~95 (ETZ, ETCZ ) | ||
Dewpoint Temp. Range (℃) | +4~+59, +4~+95 |
Heating/Cooling Rate (IEC60068-3-5) | |||||||||
LEVEL 1(K/MIN) | |||||||||
4.5 | 4.5 | 3.7 | 3.7 | 2.7 | 2.7 | 2.1 | 2.1 | 1.7 | 1.7 |
LEVEL 2(K/MIN) | |||||||||
5.5 | 5.5 | 4.5 | 4.5 | 3.2 | 3.2 | 2.6 | 2.6 | 2 | 2 |
Temp. Constancy (K) | ±0.1~ ±0.5 ( IEC60068-3-5) | |||
Temp. Homogeneity (K) | ±0.5 ~ ±3.0 ( IEC60068-3-5) | |||
Humid. Constancy Entry (%RH) | ±1.0 ~ ±5.0 ( IEC60068-3-5) (ECZ, ETCZ Series) | |||
Door Dimension Power (mm) | 900W*1900H(Single Door), 2000W*2000H (Double Door) | |||
Power | AC380V±10%,50Hz,3/N/PE |
Options | Standard Version |
☆ Independent sensors for specimen protection (NE60519-2.1993) | ☆ 1 Entry port Ø 100mm |
☆ Spare parts package | ☆ 1 Silicon stopper for entry port |
☆ E-management and cyber-software | ☆ Automatic water supply |
☆ Nitrogen gas auxiliary device |
Operation System Display
Enter Your Message