Detail Information

Product Name: Gas Test Chamber Temperature Range: 20℃~+45℃
Power Supply: AC380V ±10%, 50HZ, 3/N/PE Customized Support: OEM, ODM
Humidity Range: 50%RH~85%RH Wind Speed: ≤1.7m/s
High Light:

AC380V Noxious Gas Test Chamber

,

IEC60068 Noxious Gas Test Chamber

,

GB/T2423 Gas Test Chamber

Product Description

Noxious Gas Test Chamber From Envsin Manufacturer & Supplier

 

This Chamber can be used to test the electroplating protective layer of electrical and electronic products, electroplating products and various gold and silver products. In particular, it is used to assess the ability of the material contact parts and the protective layer of the connecting parts to resist corrosive gases, and similar protection.

GB/T2423 IEC60068 AC380V Noxious Gas Test Chamber 0

 

Technical Parameters

Product Model:EQF2330
Temperature test Temperature range:20~+45 Remark:
1. Test in accordance with GB/T2423 and IEC60068 under empty chamber condition (except indicated).

2. The test data of the water-cooled test chamber are under the conditions that the water temperature is less than 28, the water pressure is between 0.3Mpa and 0.6Mpa, and the water pressure difference between inlet and outlet is more than 0.2Mpa.
Temperature uniformity:≤2.0℃
Temperature deviation:≤±2.0℃
Temperature fluctuation:≤±0.5℃
Temperature rate:≥1℃/min
Temperature and humidity climatic test Humidity temperature control range:20℃~+50℃
Humidity range:50%RH~85%RH
Humidity uniformity:±5.0%RH
Humidity deviation: +2.0%~3%RH(>75%RH)±5%R.H(≤75%RH)
Humidity fluctuation:±5.0%RH
Gas concentration 5ppb~30ppm
The deviation of gas concentration in the chamber shall not exceed 10% of the set concentration,Concentration uniformity deviation of 10%
The empty chamber operation request meets the IEC60068-2-60 standard
Gas concentration measurement By calculating the volume ratio and then through the mass flowmeter measurement
On load capacity 500W,The temperature can be stable, the chamber body is not deformed
Noise Working noise ≤70dB (measured at 1 meter away from the test chamber)
Wind speed ≤1.7m/s
Gas flowmeter H2S,SO2,NO2,CI2

 

 Prohibitions and Instructions

 

Warnings: Below are prohibited

  • Inflammable, explosive, volatile, corrosive items
  • Living things with vital signs;
  • Particle radiation and strong electromagnetic interference;
  • Easy thermal deformation.

Load Requirements:

  • Max. 2/5 of total volume that samples occupy in test space;
  • Any surface area that faces air circulation direction takes up no more than 1/3 of the internal surface area of the Chamber. That too big a specimen will affect the air uniformity and ventilation.

Applications:

  • The chamber is suitable for testing for large size or a larger number of machine products, components, materials.It can go on high temperature, low temperature, high temperature cycle test and damp heat test. Mainly used to detect that adaptive capacity and characteristics of the product whether change when withstands climate conditions or a combination of temperature and humidity.

 

 