28m3 NE60519-2.1993 Walk In Environmental Test Chamber APP Mobility Management
Contact Person : Jack He
Phone Number : +86-13842409825
WhatsApp : +8613842409825
Minimum Order Quantity : | 1 | Price : | To Be Negotiated |
---|---|---|---|
Packaging Details : | Wooden packing | Delivery Time : | 30-35 business days after received the payment |
Payment Terms : | L/C, T/T, Western Union | Supply Ability : | 25 per month |
Place of Origin: | China | Brand Name: | Envsin |
---|---|---|---|
Certification: | CE and ISO | Model Number: | ETCZ4010 |
Detail Information |
|||
Product Name: | Environmental Test Chamber | Test Space: | 8m³ |
---|---|---|---|
Internal Size: | 2000W×2000D×2056H (mm) | External Size: | 2240Wx2240Dx3800H (mm) |
Power Supply: | AC230V/400V,50Hz±1, 60A, 3/N/PE, Earth Resistance < 4Ω | Max. Power: | 33KW |
Max. Currency: | 60A | Cooling Method: | Water-Cooled(Cooling System) |
High Light: | 8m3 Environmental Test Chamber,230V Walk In Test Chamber,CE Walk In Test Chamber |
Product Description
8m³ Environmental Test Chamber Water-Cooled Walk In Environmental Test Room
Designed for high and low temperature test and rapid temperature change ESS test for whole product, machine component, and materials.
This machine can test samples of or not of thermal dispersion. For samples of thermal dispersion, the thermal dispersion power should be no more than the refrigerating power of the test chamber. The thermal conditioning process is dynamic, and the refrigerating power varies in different temperature. Therefore if the samples are of easy thermal dispersion, please refer to the load thermal capacity graph blow for a correct test performance.
Test Space |
8m³ |
|
Internal size |
2000W×2000D×2056H (mm) |
|
External size |
2240Wx2240Dx3800H (mm) |
|
External size |
||
Power supply |
AC230V/400V,50Hz±1, 60A, 3/N/PE, earth resistance < 4Ω |
|
Max. Power |
33KW |
|
Max. Currency |
60A |
|
Cooling Method |
Water-Cooled(Cooling system) |
|
Installation Site |
|
Temp.&Humid. Parm. |
Temp. Range:-40℃~+100℃ |
Note:
1.Test with no load, meeting GB/T5170-2008, if no other special notice.
2.Ambient Temp 25℃, good ventilation condition: function well if 35℃≥RT>25℃;Cool down rate and thermal load capacity will be affected by ambient low temperature.
3.Water-cooled water temp ≤28℃, water pressure is 0.3MPa~0.6Mpa, water pressure difference is ≥0.2Mpa. |
Temp. Uniformity:±0.5℃~±2.0℃ | ||
Temp Deviation:±0.1℃~±2.0℃ | ||
Temp Fluctuation:±0.1℃~±0.5℃ | ||
Temp Gradient:<2.0℃ | ||
Heat up Rate:4K/min | ||
Cool down Rate:4K/min | ||
Climate Test
|
Temp. Range:+10℃~+95℃ | |
Humid. Range: 10%RH~98%RH | ||
Humid. Fluctuation: ±1.0℃~±3.0℃ | ||
Temp. Uniformity: ±0.5℃~±2.0℃ | ||
Humid. Uniformity: ±1.0%RH~±5.0%RH | ||
Temp. Deviation: ±0.1℃~±2.0℃ | ||
Humid. Deviation: ±1.0%RH~±5.0%RH | ||
Temp. Fluctuation: ±0.1℃~±0.5℃ | ||
Humid.Fluctuation:±1.0%RH~±3.0%RH | ||
Temp. Gradient: <2.0℃ | ||
Humid. Gradient: <5.0%RH | ||
Test Standards |
1. GB/T 2423.1 Low Temp. Test Method 2. GB/T 2423.2 High Temp. Test Method 3. GB/T2423.34 Temp. and Humid Cycling Test Method 4. IEC60068-2 Temp. and Humid Test Method 5. GJB 150.3 High Temp. Test Method 6. GJB 150.4 Low Temp. Test Method 7. GJB 150.9 Temp. and Humid Cycling Test Method 8. MIL-STD-202G-103B Humid. Test Method |
|
Load thermal capacity graph (calorific value of specimen) | In temperature test, the thermal capacity for loading varies from different temperature value. | |
Temperature and Humidity Control Chart |